Noninvasive Ambulatory Hemodynamic Monitoring Based on Electrocardiogram and Impedance Cardiography
DOI:
https://doi.org/10.3993/jfbim00162Keywords:
Electrocardiogram;Impedance Cardiography;Noninvasive Ambulatory Hemodynamic Monitoring;Characteristic Points;Cardiac Output;Cardiac Index;Postural ChangeAbstract
Two different noninvasive ambulatory hemodynamic monitoring devices (belt-type and chest-type) based\r on Electrocardiogram (ECG) and Impedance Cardiography (ICG) has been designed, fabricated and\r tested. Hardware designs of the two ambulatory hemodynamic monitoring devices based on Tetrapolar\r configuration and placement of the electrodes were introduced. Hemodynamic parameters such as Stroke\r Volume (SV), Cardiac Output (CO) and Cardiac Index (CI) can be estimated according to Kubicek\r Formula by extracting characteristic points and characteristic periods in the ICG Signal. CI is CO\r divided by body surface area. CI is comparable among different people because it is not influenced\r by the body height and weight. The body's physiological responses to postural change are primarily a\r reaction to the change of gravity, because change of gravity can cause changes of blood flow distribution\r in organs and tissues. Different body postures (upright or supine) can result in reducing or rising of\r CO. The detection accuracy has also been validated with a commercial ICG detection instrument from\r Mindray company, and the correlation coefficient is 0.83.Downloads